Features
■ High power: AC 5kV / 100mA / 500VA output
■ High security:
High-voltage floating output design, in line with the safety
requirements of EU standards EN50191 (only TH9110)
Electric shock protection function
■ High resolution: 7 inch 800 × 480 dots, TFT-LCD display
■ Brand-new operation interface, Chinese and English menu
■ ARC detection function
■ Contact check function (OSC)
■ Breakdown voltage test function
■ One-key screen capture function
■ One-key recording function
■ Rear panel output function to facilitate automated production line testing
■ Storage: 100 files, up to 50 steps per file
Model | TH9110 | TH9110A | |
Withstand Voltage Test | |||
Output Voltage | AC | 0.05 – 5kV Load Variance:1% Accuracy:1% Resolution: 2V | |
DC | 0.05 – 6kV Load Variance:1% Accuracy:1% Resolution: 2V | ||
Current Test Range | AC | 0.001mA – 120mA(Voltage≤4kV);0.001mA – 100mA(Voltage>4kV) Accuracy:1% Resolution:1μA | |
DC | 0.0001mA – 25mA Accuracy:1% Resolution:0.1 μA | ||
Output Power | 500VA | ||
Insulation Resistance Test | |||
Output Voltage | : 0.05 – 5kV Resolution:2V Accuracy:1% of set value + 0.1% full scale | ||
Resistance Test Range | 1MΩ-50.0GΩ Resolution:0.1MΩ | ||
Discharge Function | Automatic discharge after the end of the test | ||
ARC Detection | AC | 1mA – 20mA | |
DC | 1mA – 10mA | ||
Contact Check Function | OSC open and short:600Hz, 0.1s | ||
Security Features | |||
High Voltage Floating Output | Leakage current <3 mA | —— | |
Electric Shock Protection | 0.5mA ±0.25mA | ||
Other Protection | Start protection, panel operation password protection | ||
Alarm Indication | PASS: short tone, green light; FAIL: long tone, red light | ||
Memory | groups, 50 steps per group | ||
General Parameters | |||
Voltage Rise Time | 0.1s — 999.9s | ||
Test Time Setting(AC/DC) | 0.3s — 999s | ||
Voltage Fall Time | 0.1s — 999.9s | ||
Waiting Time (IR) | 0.2s — 999.9s | ||
Time Accuracy | ±(1%+0.1s) |