Test voltage DC5000V/2500V/1000V/500V/250V
Insulation Resistance up to 1TΩ
Short circuit current up to 4mA
Dielectric Absorption Ratio(DAR)
Polarisation Index(PI)
Auto discharge function
Data hold(Auto)
Auto power save
Specifications
Test Voltage(DC) | 250V | 500V | 1000V | 2500V | 5000V | ||
---|---|---|---|---|---|---|---|
Range | 0.0~104.9Ω | 0.0~99.9MΩ 80~1049MΩ | 0.0~99.9MΩ 80~999MΩ 0.80~2.09GΩ | 0.0~99.9MΩ 80~999MΩ 0.80~9.99GΩ 8.0~104.9GΩ | 0.0~99.9MΩ 80~999MΩ 0.80~9.99GΩ 8.0~99.9GΩ | 80~1000GΩ | 1001~1199GΩ |
Accuracy | ±5%+3 | ±5%+3 | ±5%+3 | ±5%+3 | ±5%+3 | ±20% | – |
Open circuit voltage | DC250V 0%~+20% | DC500V 0%~+20% | DC1000V 0%~+20% | DC2500V 0%~+20% | DC5000V 0%~+20% |
Rated current | 3.0mA±0.5mA |
Short-circuit current | 3mA~4mA |
Voltage measurement | AC:30~1000V(50/60Hz), DC:30~1000V, Accuracy:±(2%+3dgt) |
Battery monitor | 4-step indication |
IP rate | IP54 |
Battery | LR14(C alkaline battery)1.5V×8 |
Size/Mass | H188×W225×D97mm/Approx. 1750g |
Safety standard | IEC61010 CAT.IV 600V |