Benefits
- Safety concept meets the DIN EN 1539-2010-08 standard
- Short drying time with high temperature accuracy and defined ventilation
- Silicone free and dust-free inner chamber
- Replaceable fresh air cartridges are easy to change and ensure constant fresh air quality
Important characteristics
- Temperature range: +10 °C above ambient temperature to +300 °C
- APT.line™ preheating chamber technology
- Controller with time-segment and real-time programming
- 2 chrome-plated racks
- Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
- Replaceable filter cartridge, Class M6
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- Computer interface: RS 422
- Standard door gasket made from solvent-resistant FKM for use at up to 200 °C
- Optional door gasket made from silicone with limited solvent resistance for use at up to 300 °C
Specifications:
Designation | FDL115-230V¹ |
---|---|
Option model | Standard |
Order number | 9010-0292 |
Data | |
Serie | FDL |
Performance Data Temperature | |
Heating-up time to 150 °C [min] | 20 |
Recovery time after 30 seconds door open at 150 °C [min] | 12 |
Temperature fluctuation [± K] | 0.8 |
Temperature range +10 °C above ambient temperature to [°C] | 300 |
Temperature variation at 150 °C [± K] | 2.5 |
Air change data | |
Air circulation (approx.) [x/min] | 20 |
Volumetric flow rate of exhaust air acc. to EN 1539 at 50 °C [ca. L/min] | 400 |
Solvent quantity | |
Highest permitted solvent quantity (at T-180 °C, M-100g/mol, U-40g/m3, K=0,5) [g] | 6.65 |
Electrical data | |
Nominal power [kW] | 2.9 |
Phase (Nominal voltage) | 1~ |
Power frequency [Hz] | 50/60 |
Rated Voltage [V] | 230 |
Housing dimensions not incl. fittings and connections | |
Depth net [mm] | 685 |
Height net [mm] | 805 |
Width net [mm] | 830 |
Internal Dimensions | |
Depth [mm] | 435 |
Height [mm] | 435 |
Width [mm] | 600 |
Measures | |
Interior volume [L] | 115 |
Load per rack [kg] | 30 |
Net weight of the unit (empty) [kg] | 90 |
Permitted load [kg] | 60 |
Wall clearance back [mm] | 160 |
Wall clearance sidewise [mm] | 100 |
Environment-specific data | |
Energy consumption at 150 °C [Wh/h] | 1200 |
Sound-pressure level [dB(A)] | 57 |
Fixtures | |
Number of shelves (std./max.) | 2/5 |
- All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.