Code | Description |
---|---|
Nil(blank) | Without temperature input |
01 | With temperature input |
Basic Specifications
Measurement ranges | 30 mΩ, 300 mΩ, 3Ω 3 ranges |
---|---|
Accuracy | |
Measuring frequency | AC 1 kHz ± 20 Hz |
Measurement current | 7.4 mA(30 mΩ range), 1 mA(300 mΩ), 100 μA(3 Ω) |
Max. Applied voltage | 200 V DC for all ranges |
Sampling time | 2/sec (SLOW mode) 10/sec. (FAST mode) |
Response speed | 1.6 s (SLOW mode) 0.667 s (FAST mode) |
Measurement method | 4-terminal method |
Temperature measuring range | K type thermocouple: from -100.0 to 1300.0 °C J type thermocouple: from -140.0 to 1200.0 °C T type thermocouple: from -200.0 to 400.0 °C |
Power supply | LR6 (AA) Alkaline batteries × 6, AC adaptor (100/ 200 V +/- 10%) |
Weight | Approx. 1 kg |
Dimensions | 206 (W) x 65 (H) x 169 (D) mm |
Accessories | Kelvin clip , AC adaptor, LR6(AA)Alkaline batteries × 6 |