Rational-DMIS software package
Rational-DMIS is the newly developed large-scale comprehensive 3D measurement software package. In addition to the development and design of the software, the beginner’s requirements for easy learning and easy use of the software and the high-level user’s functional requirements for the software are taken into account. More importantly, it has become a model of CMM measurement software standardization.
The program of Rational-DMIS adopts the design of Full and Native DMIS until DMIS 5.0, which is completely different from the so-called “support DMIS” (the kernel is not DMIS). As we all know, DMIS has become an ISO standard (ISO 22093:2003)
UCC S3 controller
- Independent CPU processor, 32-bit high processing capacity, real-time control
- 19-inch chassis, weight 8 kg
- Directly connected to 220 volts 50/60 Hz, single phase, connected load 800 watts
- Integrated line filter, in line with EN50082, EN50081 regulations
- Temperature range operation: 10 degrees Celsius to 45 degrees Celsius
- Relative humidity: 10% to 95% non-condensing
- Performance: The fastest positioning speed with a resolution of 1 micron is 5 m/s
- Forward-looking control, acceleration distribution graph, S-distribution graph
- “Rectangle with corners and arcs” interpolation feedback system, incremental RS422, 1/T sub-count interpolation
- Counting frequency, max. 10 MHz
- Incremental sensor connection facility (5V-TTL) safety equipment/diagnosis/setting, integrated monitoring
- Multi-element light-emitting diodes, used to display the operating status of all important components on the panel
Specification
Model | JT-565 |
Measuring Range | x=510mm, y=610mm, z=508mm |
Max Load Weight | 900kg |
Net Weight | 1950kg |
Linear Scale Resolution | 0.5㎛ |
Max 3D moving speed | 500mm/s |
Max 3D Acceleration: | 700mm/s² |
Length Accuracy | ≤ 1.7+L/350 μm |
Detecting ball accuracy | ≤ 2.3μm |
Dimension | 1250*1400*2600mm |
Average Room Temperature | 18 – 22°C |
Maximum Thermal Gradients | 2 °C/day, 1°C/hour, 1°C/meter surrounding the machine. |
Humidity | 20% to 70% |
Minimum supply pressure | 5.5bar(0.55MPa) |
Air consumption | 150 NL/min |
Power | 220V ± 10%,50/60 Hz,15 A,1100W |
Probe | MH20i |