Description:
■ TH2883 series products are newly developed impulse winding testers by Tonghui. This product line makes Tonghui as the first provider of impulse winding tester from low voltage of 30V to high voltage of 10kV, single channel to multichannel (Max.:8 channels) in this industry. The instrument adopts popular 32 bit CPU and high density SMD technology, 65k color 7-inch TFT wide display screen, bringing ease for your eyes and convenience to your operation. The minimum impulse voltage of 30V, maximum impulse voltage output of 10kV, winding test of 1uH inductance value, sampling rate of 200Msps, memory depth of 6k bytes makes your test accurately. The usage of standard sample average, application of demagnetized impulse, high bandwidth analog acquisition circuit, technology of high-fidelity corona extraction as well as the opening of non-destructive test reflect the design philosophy “customer-oriented, share the future technology with you” of Tonghui.
According to the output voltage, TH2883 series is consist of 3 models:TH2883-1, TH2883-5 and TH2883-10. With minimum impulse voltage of 30V and maximum impulse voltage of 1200V, TH2883-1 low inductance impulse winding tester can test windings of 1uH low inductance value. The instrument is the ideal test product for inductance coils used by switching power supply. With impulse voltage of 100V~5000V, TH2883-5 is a standard product for testing all kinds of coils. With maximum impulse output voltage of 10kV, TH2883-10 is appropriate for interturn test of higher insulation and voltage resistance.
Standard-equipped USB Host, RS232C, USB Device and LAN interface of TH2883 series product are convenient for your fast storage of graphs and remote control.
■ With high-fidelity corona extraction algorithm (patent technology) and high bandwidth analog acquisition circuit, TH2883 series products can fully recover the corona waveform of high-frequency and makes you know more about the insulating property of products.
Technical Parameter
Model | TH2883-1 | TH2883-5 | TH2883-10 | |
Impulse Voltage | 30V-1200V, 5V steps | 100V-5000V, 10V steps | 500V-10kV, 20V steps | |
Voltage Accuracy | ±(5% set value +5V) | ±(5% set value +15V) | ±(5% set value +25V) | |
Readback Accuracy | ±(5% actual value +5V) | ±(5% actual value +15V) | ±(5% actual value +25V) | |
Channels | 1 | |||
Inductance Test Range | ≥1uH | ≥10uH | ≥20uH | |
Impulse Energy | Max.: 0.02 Joule | Max.: 0.25 Joule | Max.: 0.5 Joule | |
Test Speed | 6 times/second | 6 times/second | 3 times/second (when 10kV impulse voltage is output) | |
Pulses Applied | Max.: 32 | |||
Input Impedance | 5MΩ | |||
Display | 800×480 dots, 65k color TFT; Waveform Display Range: 600×256 | |||
Waveform Acquisition | Sampling rate: Max. 200Msps, 8 levels adjustableResolution: 8 BitsMemory Depth: 6k BytesAverage: 1 to 32 | |||
Comparison Methods | Comparison with Standard Waveform:
| |||
Waveform Measurement | Voltage/Frequency/Time | |||
Trigger Mode | Manual/External/Bus/Internal | |||
Detection Output | OK/NG display/LED/ Alarm | |||
Measurement Statistics | Statistics for measurement results | |||
Memory | 20 groups of standard waveform data and instrument setup can be stored in internal non-volatile memory.USB flash memory can be used as external memory. | |||
Interface | Handler, RS232C, USB Device, USB Host, LAN | |||
Power Supply | ||||
Power Supply | 220V ±10% 50Hz/60Hz ±5% | |||
Power Consumption | ≤200VA | |||
General Conditions | ||||
Working Environment | Temperature | 0℃ – 40℃ | ||
Humidity | ≤75% R.H. | |||
Safety and Electromagnetic Compatibility | IEC61010-1:2001,IEC61326-2-1:2005 |