The AN9651TH(F) is a nine-in-one functions (ground bond resistance/insulation resistance/AC withstand voltage/DC withstand voltage/leakage current/power/start/wait/arc detection) electrical safety comprehensive tester independently developed and produced by Ainuo Instruments, three-phase/single-phase compatible, suitable for safety performance testing of production lines, laboratories and quality inspection departments.
From medical devices to appliances, AINUO AN Series all-in-one design includes the test functionality you need to ensure your product is compliant with the most rigorous international electrical safety test standards.
The AC Hipot Test allows you to stress your product’s insulation with high voltage to ensure it will limit current flow should a fault occur.
Output 100 mA of AC Hipot/Dielectric Withstand test current. Ainuo AN Series meets the 200 mA short circuit requirement for 500 VA Hipot Testers.
The DC Hipot Test allows you to stress your product’s insulation and limit leakage current output by running a DC dielectric withstand test.
The Ground Bond: Verify the integrity of your product’s protective earth ground conductor to prevent electric shock should a fault occur.
The The Ground Continuity: Verify the presence of your product’s protective earth ground conductor to prevent electric shock should a fault occur.
The Insulation Resistance: Determine the total resistance of your product’s insulation.
The Leakage Current: Ensure your product’s leakage current remains at acceptable levels during 8 different fault conditions.
The Functional Run: Verify that your product is working properly before shipment.
Features:
Comply with CCC, IEC, EN,VDE, BS, UL,JIS standard requirements.
Latest international standard: 500VA withstand (AC 5kV/100mA), short-circuit output current exceeds 200mA. Built-in replacement leakage current test network MD card, meeting requirements of different industrial standards for different MD.
9 full tests: AC withstand, DC withstand voltage, insulation resistance, ground resistance, leakage current, operating power, low-voltage start, arc detection, delay waiting.
Intelligent test: 15 groups of test memories, 8-step programming for each group; a key to start automatically and complete all tests. Built-in isolated power unit, automatic control of supply voltage of tested products; external variable frequency power online test. Product bar code reading and identification, automatically call the test group according to the model/specifications of the products tested.
Auto control: RS232/RS485/GPIB/Ethernet interface (four choose one), bar code scanner interface, inverter power control interface, switch control interface, easily forming automation system.
Information management: optional industrial PC and software (ESRS), sharing test data with production information systems such as ERP and MEC, realizing test process and data management, improving product quality and production management level.
Simplified operation: 10.4” TFT color display, menu operation, Chinese/English, socket test tooling.
Specification of AN9651TH(F):
Features | Technical indicator | |||
AC voltage withstand test | Rated output capacity | 500VA (5000V/100mA),short-circuit current more than 200mA | ||
Output voltage | Range | static state:500~5000V;dynamic:500~2500V | ||
Resolution /Accuracy | 50V;±(2% × setting value + 5V) | |||
Output frequency | Range/Accuracy | 50Hz or 60Hz, ±(0.1% × setting) | ||
Alarm current setting | Upper limit range/Accuracy | 0.10~99.99mA, ±(2% × reading + 2 counts) | ||
Lowwer limit range/Accuracy | 0.00~99.99mA, ±(2% × reading + 2 counts) | |||
Compensation current setting | Range | 0.00~5.00mA, Automatic measurement, Compensation can be turned on or off | ||
Initial voltage setting | Range | (0%~99%) of testing voltage | ||
Time setting | Range of test time | 0.5~999.9s (Less than 1s has an impact on the accuracy of the test),error: ±(0.1%×setting+2counts) | ||
Range of ramp-up time | 0.1~999.9s,error: ±(0.1%×setting+2counts) | |||
Range of ramp-down time | 0.0~999.9s (The timing starts when the test pass),error: ±(0.1%×setting+2counts) | |||
DC voltage withstand test | Rated output capacity | 60VA(6000V/10mA) | ||
Output voltage | Range | static state:500~6000V;dynamic:500~2500V | ||
Resolution /Accuracy | 50V,±(2% × setting + 5V) | |||
Alarm current setting | Upper limit range/Accuracy | 50~9999µA, ±(2% × reading + 2 counts) | ||
Lowwer limit range/Accuracy | 0~9999µA, ±(2% × reading + 2 counts) | |||
Compensation current setting | Range | 0~500µA, Automatic measurement, Compensation can be turned on or off | ||
Initial voltage setting | Range | (0%~99%) × setting of output voltage | ||
Time setting | Range of test time | 0.5~999.9s (Less than 1s has an impact on the accuracy of the test),error: ±(0.1%×setting+2counts) | ||
Range of ramp-up time | 0.1~999.9s,error: ±(0.1%×setting+2counts) | |||
Delay decision time | 0.3~10.0s (Starting from the ramp-up time),error: ±(0.1%×setting+2counts) | |||
Arc detection | 1~9 (9 corresponds to most sensitive), 0 is Off | |||
Insulation resistance test | Rated output capacity | 1VA(1000V/1mA) | ||
Output voltage | Range/Accuracy | DC 100V-1000V,±(2% × setting + 2V) | ||
Alarm resistance setting | Range of upper limit | 1.0~99.9MΩ,100~2000MΩ(LLLL is infinite) | ||
Range of lower limit | 1.0~99.9MΩ,100~2000MΩ | |||
Time setting | Range of test time | 0.5~999.9s (Less than 1s has an impact on the accuracy of the test),error: ±(0.1%×setting+2counts) | ||
Range of ramp-up time | 0.1~999.9s,error: ±(0.1%×setting+2counts) | |||
Delay decision time | 0.3~99.0s (Starting from the ramp-up time),error: ±(0.1%×setting+2counts) | |||
Insulation resistance measurement | Range | 1.0~99.9MΩ,100~2000MΩ | ||
Accuracy | ≤200MΩ:±(2% × reading + 2 counts);>200MΩ:±(5% × reading + 5 counts) | |||
Grounding resistance test | Open circuit output voltage | <12V | ||
Output current | Range/Accuracy | 3~32A,±(2% × setting + 0.2A) | ||
Current frequency | Range/Accuracy | 50Hz or 60Hz,±0.1% × setting | ||
Alarm limit | Range of resistance upper limit / Accuracy | 3A≤Output current≤10A:10~600Mω; ±(2% × reading + 2 counts) 11A≤Output current≤25A: 10~300mΩ; ±(2% × reading + 2 counts) 26A≤Output current≤32A: 10~200mΩ; ±(2% × reading + 2 counts) | ||
Range of resistance lower limit / Accuracy | 0~100mΩ; ±(2% × reading + 2 counts) | |||
Compensation resistance | Range | 0~200mΩ,automatic testing, compensation function can be set to ON/OFF | ||
Testing time | Range | 0.5~999.9s (Less than 1s has an impact on the accuracy of the test),error: ±(0.1%×setting+2counts) | ||
Leakage Current Test | Test type setting | Single phase/3 phase load, static/dynamic leakage test | ||
MD network selection | Standard GB/T12113 figure 4 network; other test network optional | |||
Leakage current alarm setting | Upper current range | 50~9999µA | ||
Lower current range | 0~5000µA | |||
Leakage current measurement | range | 50~9999µA | ||
Accuracy | DC~10kHz:±(2% × reading + 2 counts); 10k~1MHz:±(5% × reading + 5 counts); | |||
Output voltage | Range/Accuracy | 50.0~280.0V(phase voltage), ±(0.5% × reading + 2 counts) | ||
Testing time | range/Accuracy | 1.0 ~999.9s, error: ±(0.1%×setting+2counts) | ||
Run Power Run | Alarm limit setting | Upper power range | single-phase:20~6000W;Three-phases:60~20000W | |
Lower power range | single-phase: 0~6000W;Three-phases: 0~20000W | |||
Test time setting | Range | 0.5~999.9s (Less than 1s has an impact on the accuracy of the test) | ||
voltage measurement | range/Accuracy | 50.0~280.0V(phase voltage), ±(0.5% × reading + 2 counts) | ||
current measurement | range/Accuracy | 0.30~30.00A,±(0.5% × reading + 2 counts) | ||
Active power measurement | range/ Accuracy | 20.0~999.9W,1000~9999W,10.00~20.00kW;±(0.5%×reading value +2 counts) | ||
Low voltage starting test | Alarm limit setting | Upper current range | 0.30~30.00A | |
Lower current range | 0 ~30.00A | |||
Testing time | Range | 0.5~999.9s (Less than 1s has an impact on the accuracy of the test) | ||
voltage measurement | range / Accuracy | 50.0~280.0V (phase voltage), ±(0.5% × reading + 2 counts) | ||
current measurement | range / Accuracy | 0.30~30.00A,±(0.5% × reading + 2 counts) | ||
DUT power control | Isolation transformer input voltage, frequency | 0~300V(phase voltage),45~65Hz | ||
Isolation transformer output voltage ratio | Leakage current test | 1.06 times of rated voltage | ||
Running power test | 1.00 times of rated voltage | |||
Low voltage start test | 0.85 times of rated voltage | |||
Isolation transformer capacity | Single phase load: 6kVA,three phase load: 20kVA; 45kW/60kW/90Kw is optional. | |||
Delay time | Delay time setting | Range/Accuracy | 1.0~999.9s,LLLL is infinite, ±(0.1% × setting + 2counts) | |
Load power mode setting | During delay, load power can be set to ON/OFF | |||
Interface | Remote interface | Standard,manual remote control box(start/stop) | ||
Alarm lamp interface | Standard, 3 color lamp(test/pass/fail) | |||
Communication interface | Standard RS232,internet、RS485、GPIB is optional | |||
Barcode scanner interface | Standard, Can work with barcode scanner to automatically identify barcode information | |||
AC power supply control interface | Standard,combine with AN97 series frequency conversion power | |||
PLC interface | Standard,start/stop/test/pass/fail output | |||
Printer interface | Standard |