INSTRUCTIONS ZX1379 Series
ZX1379M DC bias current source can be easily connected to an external LCR meter through a standard RS232C interface. Up to 8 ZX1379S can be stacked, with an output current of 225A and an output voltage of up to 11V. ZX1379MX is a bridge bias current all-in-one machine with a built-in high-frequency LCR digital bridge and a 0-25A DC bias current source, and can also be connected to ZX1379S to expand the current. ZX1379S is a single 0-25A DC bias current source, S stands for slave, and needs to be connected to ZX1379M or ZX1379MX to use.
PERFORMANCE
- Provide single 0-25A constant current output
- Support up to 8 slaves, maximum 225A constant current output
- Overload alarm indication and output cut-off
- 0.25mA fine current step
- 0Hz-2MHz frequency response
- Two current output modes: single current and step scanning
- Graphical operation, Chinese and English interface
- Can be controlled by external LCR meter via RS232C interface using SCPI commands
Technical parameters
Parameter | ZX1379M | ZX1379MX | ZX1379S |
---|---|---|---|
Display | 7″ LCD | 7″ LCD | LED indicator light |
Operation | Physical buttons + Foot switch | Physical buttons + Foot switch | Controlled by the host |
Supported Measurement Frequency | 0Hz-2MHz | 0Hz-2MHz | Not available |
Current Range | 0-25A | 0-25A | Single unit 25A, expandable to 8 units, total 225A |
Range | 0mA-1.000A | 1.000A-5.000A | 5.0A-225A |
Step | 0.25mA | 1.25mA | 6.25mA*N (number of linked units) |
Scan Adjustment Time | 4ms-3600s | 10ms-3600s | 20ms-3600s |
Accuracy | ± (1% + 5A) | ±2% | ±3% |
Maximum Output Voltage | 11V (11V can be output under all current conditions!!!) | 11V | 11V |
Maximum Allowable DC Resistance | Rmax = Vmax/I (Ω) (Refer to user manual for Rmax calculation) | Rmax = Vmax/I (Ω) | Rmax = Vmax/I (Ω) |
Maximum Allowable Inductance | Lmax = Vmax/(di/dt) (mH) (Refer to user manual for Lmax calculation) | Lmax = Vmax/(di/dt) (mH) | Lmax = Vmax/(di/dt) (mH) |
Range Selection Method | Automatic | ||
Start and Stop Control Method | Physical START/STOP buttons, 4 types of foot switches, bus control | ||
Continuous Maximum Current Loading Time | Conservative 2-3 hours, uninterrupted output | ||
Functions | Device self-test; 99 custom file management; English-Chinese bilingual; SCPI command set | ||
Interface | RS232, SlaverLink | ||
Dimensions (WxHxD) | 430mm x 178mm x 465mm | ||
Weight (Net) | 19kg |