Constant climate chamber HPP260eco
Standard delivery Constant climate chamber
On this page, you can find all the essential technical data on the Memmert stability chamber HPP. Our customer relations team will be pleased to help if you want further information
Temperature
Setting temperature range | without light, with humidity: +5 to +70 °C |
---|---|
Setting temperature range | with light, with humdity: +5 to +70 °C |
Setting temperature range | with light, without humditiy: 0 to +70 °C |
Setting temperature range | without light, without humidity: 0 to +70 °C |
Working temperature range | with light, without or with humidity: 15 to +40 °C |
Working temperature range | without light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C |
Working temperature range | without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C |
Setting accuracy temperature | 0.1 °C |
Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Humidity
Setting range humidity | 10 – 90 % rh |
---|---|
Humidity | humidity supply with distilled water from external tank by self-priming pump |
Humidification | humidification by hot steam generator |
Dehumidification | dehumidification by cold trap using the Peltier technology |
Setting accuracy humidity | 0.5 % rh |
Control technology
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
adjustable parameters | temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function HeatBALANCE | adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable values each, temperature and humidity |
Standard equipment
Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
---|---|
Internals | 2 stainless steel grid(s), electropolished |
Standard accessories | Water tank including connection hose (110-750: 2,5 liter, 1060-2200: 10 liter) |
Works calibration certificate | standard values +10 °C and +37 °C, 60 % rh at +30 °C |
Door | inner glass door |
Stainless steel interior
Dimensions | w(A) x h(B) x d(C): 640 x 800 x 500 mm (d less 32 mm for fan – Peltier) |
---|---|
Volume | 256 l |
Max. number of internals | 9 |
Max. loading of chamber | 200 kg |
Max. loading per internal | 20 kg |
Textured stainless steel casing
Dimensions | w(D) x h(E) x d(F): 824 x 1183 x 655 mm (d +56mm door handle & +111mm Peltier element) |
---|---|
Housing | rear zinc-plated steel |