Description
Best heating performance combined with high cooling capacity – those are the key features of the W91 systems. These models are just as ready for embedding into pilot plants as they are for use in material and component testing.
Feature
- Rapid heating and cooling
- Wide working temperature ranges using one bath fluid
- Designed to quickly compensate exothermic and endothermic reactions
- Unsurpassed power and efficiency
- Heating capacity up to 36 kW
- Cooling capacity up to 11 kW
- Pump pressure up to 5.5 bar, max. flow rate 80 l/min
- Temperature stability ±0.05 °C … ±0.2 °C
- 5,7” industrial color TFT touch screen
- Connections for USB, Ethernet, RS232, and Alarm output
- Modbus
- Ambient temperature range +5 °C to +40 °C
- Space optimized design creates more room directly next to the units
- Extensive warning, protection, and monitoring functions with detailed self-explanatory messages
- ICC cascade control for extraordinary precision, temperature stability ±0.05 °C … ±0.2 °C
- Electronic level indication
- Powerful circulating pumps, select a stage or a specific pressure
- SD-Card slot
- Analog connections, RS485, Profibus DP (accessory)
- Intelligent warning and safety functions
Specification
orking temperature range-91 … +250 °CTemperature stability±0.05 … ±0.2 °CHeating capacity36 kWCooling capacity
Refrigerant variant | ||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Order No. | 9421913.07.TT | |||||||||||||||||||||||||||||||||
|
Refrigerant variant S1 | ||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Order No. | 9421913.S1.07.TT | |||||||||||||||||||||||||||||||||
|
Pump capacity flow rate18 … 70 l/minPump capacity flow pressure0.8 … 5.5 barDimensions
(W × L × H)95 x 127 x 190 cm