- The 6913 motor stator test system collocates the MT-6913 and the 7605 tester. Optional 6365 benchtop LCR Meter for testing inductor. It is provided with eight 4-line independent measurement test points and 13 channels. The testing projects include the DC resistance tests which may detect the errors of copper wires that is over-tight during winding, broken or connecting. Besides, the Hipot testing on the required safety regulations, and the client respect project of the Impulse/Surge are included as well. With the most advanced Impulse/Surge tester takes waveform comparison to avoid the Corona Phenomenon that is caused by possible damage. It provides a complete solution of standard test project to the international manufacturers and serves as a further checker for your products. The motor stator test systems that match up the programmable interface and the simple operation can be quickly imported into the production lines and the quality control center.
Standard system is the Super Space-Saving Desktops Tester.
Optional system is the Systematic Modular Cabinet Tester.
Feature
- ‒ One stop motor coil measurement solution, measuring items included
inductance, AC/DC hipot, impulse winding, DCR (temperature
compensation), polarity
‒ 4-wire DCR measurement
‒ The 691x impulse winding tester is the host instrument, all settings can
be completed on the host
‒ Storage file up to 100 sets
‒ It can test multiple sets of DUTs simultaneously (depending on the
number of winding set of the product)
‒ Arcing detection function
‒ The impulse winding tester with waveform comparison is a nondestructive analysis
‒ PC connection data analysis software is available
Application
- Motor, various stator and rotor coils
Measurement Data
6913 DCR | |||
Range | 0.1mΩ ~ 100KΩ (4-wire measurement) | ||
Accuracy | Low Resistance:0.1mΩ ~ 1Ω ±( 0.2% ± 1mΩ) High Resistance:1Ω ~ 100KΩ ±0.1% | ||
6913L DCR | |||
Range | 0.1mΩ ~ 150Ω (4-wire measurement) | ||
Accuracy | 0.1mΩ ~ 200mΩ ±( 0.1% ± 0.5mΩ) , 200mΩ ~ 1Ω ±( 0.1% ± 2mΩ), 1Ω ~ 150Ω ±0.5% | ||
Hipot | |||
Voltage | AC 0.1KV ~ 3KV DC 0.1KV ~ 3KV | ||
Leakage Current | AC 0 ~ 10mA DC 0 ~ 5mA | ||
Arcing | 0 ~ 9 | ||
Measurement Time | 0.1 ~ 99.9 Continously Adjustable | ||
Accuracy | 3% | ||
IR | |||
Voltage | 100V ~ 1000V ±5% | ||
Range | 1MΩ ~ 9.9GΩ | ||
Measurement Time | 0.1 ~ 99.9 Continously Adjustable | ||
Accuracy | 3% | ||
Short Circuit | |||
Voltage | 0.2KV ~ 3KV ±2% | ||
Measurement Item | Total Area、Area size comparison、Flutter、Waveform comparison、Corona | ||
Accuracy | 2% | ||
6365/6420 LCR Meter | |||
Test Frequency | 10Hz-200kHz | ||
AC Drive Level | 10mV to 2Vrms (1m Vrms steps) |
Specification
Channel | 13 channels |
Indicator | Pass/Fail Screen Diaplay/Sound |
Fixture | According to the test requirements |
Power Supply | 90 ~ 132Vac or 198 ~ 264Vac (selectable) |
Interface | RS-232 |
Flash Memory | Storage of 100 rewritable data sets |
Operation | Manual、RS-232 |
Environment | Temperature : 10°C ~ 40°C Humidity : 20 ~ 90% |
7605 Dimension (W*H*D) / Weight | 435x145x522 mm / 14 Kg |
6913 Dimension (W*H*D) / Weight | 435x190x522 mm / 15 Kg |
6365/6420 Dimension (W*H*D) / Weight | 435x145x522 mm / 9 kg |
7605 Display | 128*64 dot-matrix display |
6913 Display | 320*240 dot-matrix display |
6365/6420 Display | 480*272 Color screen |
Parameters Measurement Ranges
Item | Range |
DCR | 0.1mΩ ~ 100KΩ |
IR | 1MΩ ~ 9.9GΩ |
Hipot Leakage Current | 0.1mA ~ 10.0mA |
L.S | Impulse Voltage 200V ~ 5000V |
Ls Lp | 0.01nH-100kH |