Performance characteristics
- Output power: AC: 5kV/10mA; DC: 6kV/5mA
- No IR test function
- 480×272 dot-matrix, TFT-LCD display
- Rapidly discharging and arc detection
- Hold 20 testing steps; 4 testing modes selectable
- Brand new operation interface and concise interface operation design
- Lock keyboard
- PLC interface
- Randomly set voltage rising time and testing time in 999.9 seconds
Technical Parameter
Model | TH9310A | |||
Withstanding Voltage Test | ||||
Output Voltage | AC | 0.05 —5.00kV ± (2% reading+5digits), (50Hz、60Hz optional) | ||
DC | 0.05 —6.00kV ± (2% reading+5digits) | |||
Voltage Adjustment Rate | ≤ (1% – 5V)(rated power) | |||
Current TestRange | AC | 0.000mA – 10.00mA ± (2% reading+2digits) | ||
DC | 0uA – 5.00mA ± (2% reading+2digits) | |||
Discharge Function | Discharge after test ends(DCW) | |||
Discharge Function | Discharge after test ends | |||
Arc Detection | ||||
Measurement Range | AC | 1 – 9 levels (factory default 5)(20mA,18mA, 16mA, 14mA, 12mA, 10mA, 7.7mA, 5.5mA, 2.8mA respectively) | ||
DC | 1 – 9 levels | |||
General Specification | ||||
Memory | 20 test files with each file can store 20 test steps | |||
Voltage Rising Time | 0.1s – 999.9s | |||
Test Time Setting(AC/DC) | 0.2s – 999.9s | |||
Voltage Drop Time | 0.1s – 999.9s | |||
Time Accuracy | ±(1%+0.1s) | |||
Standard | HANDLER, RS232, USBDRV(PC interface), USBHOST(System upgrade interface) |