CO2 incubator ICO105
Standard delivery CO2 incubator
On this page, you can find all the essential technical data on the Memmert CO2 incubator ICO. Our customer relations team will be pleased to help if you want further information
Temperature
Working-temperature range | 5 °C above ambient temperature up to +50 °C Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors) |
---|---|
Setting temperature range | +18 to +50 °C |
Setting accuracy temperature | 0.1 °C |
Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Temperature variation in chamber | at + 37 °C +/- 0.3 K |
Temperature fluctuations with time |
Humidity
Humidity control (standard) | Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 % |
---|---|
Setting accuracy humidity | 0.5 % rh |
Setting range active humidity control (with option K7) |
Control of standard components
CO2 control | Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation |
---|---|
Adjustment range CO2 | 0 to 20 % CO2 |
Setting accuracy CO2 | 0,1% |
Variation in time CO2 | +/- 0.2 % CO2 |
Adjustment range O2 | 1 to 20 % O2 |
Setting accuracy O2 | 0.1 % O2 |