Brief Introduction
TH2816A is a new precision LCR meter combined with years of technical experience and newest measurement technology of instrument industry. With powerful measurement functions, high performance and low cost, TH2816A/TH2816B/TH2817A have been one of the world advanced instruments, and it provides users a super value measurement resolution and experience. The meter offers stable 6 digit resolution, wide frequency range (200kHz for TH2816A), programmable signal level (0.01V to 2.0V), up to 30 meas/sec measurement rate, 9 measurement ranges, 30Ω or 100Ω constant output impedance and friendly operation interface. TH2816Acan be used for incoming inspection of components, quality control of product line and high accuracy laboratory use. The HANDLER, GPIB, RS232C interfaces make it easy to build an automatic component test system,
Technical Parameter
Model | TH2816A | ||
Measurement Function | |||
Test Parameter | |Z|, C, L, X, B, R, G, D, Q, θ | ||
Basic Accuracy | 0.05% | ||
Equivalent Circuit | Series and Parallel | ||
Math Function | Deviation and Percent Deviation | ||
Ranging Mode | Auto, Hold | ||
Trigger Mode | Internal, Manual, External and Bus | ||
Measuring Speed | Slow: 1.5meas/sec Med: 10meas/sec Fast: 30meas/sec | ||
Correction Function | Open, Short and Load corrections | ||
Measurement Terminal | 5 terminals | ||
Averaging Rate | 1—255 , TH2816A/TH2817A only | ||
Delay Time | 0—60sec , with step of 1ms | ||
List Sweep | List sweep for up to 4 frequencies, signal levels and DC bias levels | ||
Display Mode | Direct, ΔABS, Δ%, V/I( V/I monitor), Bin number and bin counter | ||
Display | 240×64 dot-matrix LCD display, 6-digit resolution | ||
Test Signal | |||
Signal Frequency | 50Hz to 200kHz over 12,000 points | ||
Output Impedance | 30Ω , 100Ω | ||
Test Level | 10mVrms to 2.0Vrms, 10mV steps | ||
θ(RAD) | -3.14159 — 3.14159 | ||
Δ% | -999.999% — 999.999% | ||
Comparator, memory & interface | |||
Comparator Function | Bins(BIN1 to BIN9 , OUT of bins), and additional AUX bin | ||
Memory | Control settings memory for store/recall | ||
Interface | RS-232C, HANDLER, GPIB (Optional) | ||
Measurement Display Range | |||
|Z|, R,X | 0.00001Ω — 99.9999MΩ | ||
C | 0.00001pF — 999.999mF | ||
L | 0.00001μH — 9.99999kH | ||
G,B | 0.00001μS — 999.999S | ||
D | 0.00001 — 9.99999 | ||
Q | 0.00001 — 99999.9 | ||
θ(DEG) | -179.999º — 179.999º |