Features
- Signal source frequency range: 10 Hz to 1/3/5/10/20/30/50 MHz
- Basic accuracy up to ±0.08% (typical ±0.05%)
- ALC function
- Output impedance 25Ω/100Ω, switchable
- Support meter mode and list mode, sweep mode, and equivalent circuit analysis(option) function
- Built-in DC Bias voltage ±12V
- Measurement of piezoelectric element admittance circle, and can measure DC bias characteristic of capacitance value.
- Ultra-high measuring speed < 3ms
- Open circuit/short circuit/load correct function
- Up to four component parameters can be selected in the electric meter mode. The induction and DCR values can be measured and displayed simultaneously
- Auto component classification: Comparator function and Handler BIN classification function
- Can be used with various fixtures, such as: liquid dielectric material test fixture, dielectric material test fixture and magnetic material test fixture….etc.
- Using with DC bias current test system
- Support RS-232、GPIB、Handler、LAN、USB Host/Device interfaces
- PC connection data analysis software is available
Specification
Test Frequency | 6332-1/1S | 10Hz ~ 1MHz |
6332-3/3S | 10Hz ~ 3MHz | |
6332-5/5S | 10Hz ~ 5MHz | |
6332-10/10S | 10Hz ~ 10MHz | |
6332-20/20S | 10Hz ~ 20MHz | |
6632-30/30S | 10Hz ~ 30MHz | |
6632-50/50S | 10Hz ~ 50MHz | |
* Optional Equivalent Circuit Analysis function (S model) | ||
Minimum Resolution | 100mHz, 6-bit Frequency Input | |
Accuracy | 7 ppm±100mHz | |
Basic Accuracy | ± 0.08% (typical ±0.05%) | |
AC Test Signal Level | Test Signal Voltage Level | 10mV – 2Vrms(Ro 100Ω) 10mV – 1Vrms(Ro 25Ω) |
Voltage Minimum Resolution | 1mV | |
Accuracy | ALC ON:6% * Voltage ± 2mV ALC OFF:10% * Voltage ± 2mV | |
Test Signal Current Level | 200µA – 20mArms | |
Current Minimum Resolution | 10µA | |
Accuracy | ALC ON:6% * Current ± 20µA ALC OFF:10% * Current ± 20µA | |
DC Test Signal Level | 1V (Fixed) | |
Test Time (The fastest) | < 3mS | |
Output Impedance | 25Ω/100Ω, switchable (Nominal value) | |
Parameters | |Z|、|Y|、θ、X、R、G、B、L、D、Q、DCR、C、 ESR、ε and μr | |
Bias | Can be using with DC Bias | |
Measurement Mode | Meter mode, list mode, sweep mode, and optional equivalent circuit analysis function(S model) | |
Correction | Open Circuit/ Short Circuit/Load Correction | |
Measurement Circuit | Series/Parallel | |
List mode | 50 groups of list mode setting (Each group contains up to 15 steps) | |
Equivalent Circuit Analysis | Three elements、Four elements (option) | |
DC bias voltage | -12 to +12V, 0.3% ±1.5mV, 100Hz to 30MHz | |
BIN | 9 Level | |
Comparator | ABS、ΔABS、Δ%、OFF | |
Internal Storage | 100 sets LCR setting documents, 50 groups of list mode setting | |
USB Storage | LCR setting documents, list mode setting document, BMP graphics, Sweep screen and test result data | |
Interface | RS-232、GPIB、Handler、LAN、USB Host/Device | |
Option | PC connection software | |
Equivalent Circuit Analysis | ||
Power Supply | Voltage 90V ~ 264Vac | |
Frequency 47Hz ~ 63Hz | ||
Power Consumption | Low power consumption: Maximum 30W (Nominal value) | |
Display | 7.0”TFT, 800×480 color screen | |
Environment | Temperature: 10℃ ~ 40℃ Humidity: ≦90% | |
Dimension (W*H*D) | 336x147x340 mm | |
Weight | 3.95Kg |
Parameters Measurement Ranges
|Z| | 0.000 mΩ to 9999.99 MΩ |
R, X | ± 0.000 mΩ to 9999.99 MΩ |
|Y| | 0.00000 μS to 999.999 kS |
G, B | ± 0.00000 μS to 999.999 kS |
θRAD | ± 0.00000 to 3.14159 |
θDEG | ± 0.000° to 180.000° |
Cs, Cp | ± 0.00000 pF to 9999.99 F |
Ls, Lp | ± 0.00 nH to 9999.99 kH |
D | 0.00000 to 9999.99 |
Q | 0.00 to 9999.99 |
Δ | ± 0.00% to 9999.99% |
Rdc | 0.00 mΩ to 99.9999 MΩ |
εr’ εr” | 0 to 10000 |
μr’ μr” | 0 to 10000 |
Applications
Passive Components | Capacitor, Inductor, Resistor, Transformer, Ceramic resonator, Quartz Crystal |
Semiconductor Components | The CV characteristics analysis of varactor diodes, Diodes |
Dielectric Material | Estimation on permittivity and consumption tangent of plastic, ceramic and PCB |
Electromagnetic Material | |
Other Components | Estimation of the impedance of PCB components |