Benefits
- Uniform drying conditions thanks to APT.line™ technology
- Identical test conditions throughout the chamber interior independent of sample size and quantity
- Outstanding thermal insulation saves operation costs
- Advanced programming options
Important characteristics
- Temperature range: +5 °C above ambient temperature to +300 °C
- APT.line™ preheating chamber technology
- Adjustable fan speed
- Program-controlled ventilation flap
- Color LCD controller with time-segment programming
- 2 chrome-plated racks
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- Computer interface: RS 422
Specifications:
Designation | M053-230V¹ |
---|---|
Option model | Standard |
Order number | 9010-0201 |
Data | |
Serie | M |
Performance Data Temperature | |
Heating-up time to 150 °C [min] | 15 |
Recovery time after 30 seconds door open at 150 °C [min] | 3 |
Temperature fluctuation at 150 °C [± K] | 0.3 |
Temperature range +5 °C above ambient temperature to | 300 °C |
Temperature variation at 150 °C [± K] | 1.3 |
Air change data | |
Air change (approx.) at 150 °C [x/h] | 192 |
Electrical data | |
Nominal power [kW] | 1.2 |
Phase (Nominal voltage) | 1~ |
Power frequency [Hz] | 50/60 |
Rated Voltage [V] | 230 |
Housing dimensions not incl. fittings and connections | |
Depth net [mm] | 575 |
Height net [mm] | 780 |
Width net [mm] | 635 |
Internal Dimensions | |
Depth [mm] | 340 |
Height [mm] | 400 |
Width [mm] | 400 |
Measures | |
Interior volume [L] | 53 |
Load per rack [kg] | 15 |
Net weight of the unit (empty) [kg] | 61 |
Permitted load [kg] | 40 |
Wall clearance back [mm] | 160 |
Wall clearance sidewise [mm] | 100 |
Environment-specific data | |
Energy consumption at 150 °C [Wh/h] | 300 |
Fixtures | |
Number of shelves (std./max.) | 2/5 |
- All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.