Climate chamber ICH750L
Standard delivery Climate chamber
On this page, you can find all the essential technical data on the Memmert climate chamber ICH. Our customer relations team will be pleased to help if you want further information
Temperature
Working temperature range | without humidity: 0 to +60 °C |
---|---|
Setting temperature range | 0 to +60 °C |
Working temperature range | from +10°C up to +60°C with humidity and/or light |
Setting accuracy temperature | 0.1 °C |
Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Humidity
Humidity | humidity supply with distilled water from external tank by self-priming pump |
---|---|
Humidification | humidification by hot steam generator |
Humidity | active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity – resolution of display 0.1 %, setting accuracy 0.5 % |
Control of standard components
Light | daylight and UV-light can be programmed individually via controller |
---|---|
Light | cold white light +/- 6 %, for UV-light +/- 10 % |
Light | illumination unit consisting of 4 fluorescent lights with cold white light (standard illuminant D65, 6.500K) and 2 UV-lamps in the spectral range of 320-400 nm |
Control technology
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable values each, temperature/humidity, for two freely selectable values CO2 |
adjustable parameters | temperature (Celsius or Fahrenheit), fan speed, light, programme time, time zones, summertime/wintertime |
Heating concept
Air jacket | air jacket heating system for gentle all-around heating |
---|---|
Heating concept | no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior |
Cooling | energy-saving use of CFC-free cooling/heating system (refrigerant R134a) |
Defrosting | highly efficient automatic defrosting system |
Standard equipment
Entry port | Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed by a silicone stopper, standard position at the back |
---|---|
Internals | 2 stainless steel grid(s), electropolished |
Works calibration certificate | standard values +10 °C and +37 °C, 60 % rh at +30 °C |
Door | inner glass doors |
Door | fully insulated stainless steel doors with 2-point locking (compression door lock) |
Stainless steel interior
Dimensions | w(A) x h(B) x d(C): 1040 x 1200 x 600 mm (d less 33 mm for fan) |
---|---|
Volume | 749 l |
Max. number of internals | 14 |
Max. loading of chamber | 200 kg |
Max. loading per internal | 30 kg |
Textured stainless steel casing
Dimensions | w(D) x h(E) x d(F): 1224 x 1950 x 785 mm (d +56mm door handle) |
---|---|
Installation | on lockable castors |
Housing | rear zinc-plated steel |