Descriptions
6355/6356+6243H/6243/6223 provide a frequency response range of 100Hz to 200k/500kHz and a maximum output current of 640A/320A/120A. These systems are used for testing inductors, coils, ferrites, and core saturation currents.
Features
- Frequency Response 100Hz-200k/500kHz
- Supports maximum output DC bias current of 640A/320A/120A
- DCR measurement function
- Can sustain maximum power output for long periods of time
- Output impedance: 100Ω
- Open/short circuit calibration function
- Supports meter measurement mode/multi-step list testing
| 6350+ Series | ||||||
| Model | 6355+ 6243H/6243/62236356+ 6243H/6243/6223 | |||||
| Frequency Response | 100Hz-200kHz (6355+ 6243H/6243/6223)100Hz-500kHz (6356+ 6243H/6243/6223) | |||||
| DC Bias Current Max. | 640A (6243H)320A (6243) 120A (6223) | |||||
| Power consumption Max. | 640W (6243H/6243)320W (6223) | |||||
| Selection | ||||||
| Model | 6355+6243H | 6355+6243 | 6355+6223 | 6356+6243H | 6356+6243 | 6356+6223 |
| Frequency Response | 100Hz-200kHz | 100Hz-200kHz | 100Hz-200kHz | 100Hz-500kHz | 100Hz-500kHz | 100Hz-500kHz |
| Output Current Max. | 640A | 320A | 120A | 640A | 320A | 120A |
| Power consumption | 640W | 640W | 320W | 640W | 640W | 320W |
| units of the DC Bias | 16 | 8 | 6 | 16 | 8 | 6 |
| Output Current (unit) | 40A | 40A | 20A | 40A | 40A | 20A |
SPEC
| Current Accuracy | 0.000A-1.000A 1%+5mA |
| 1.001A-5.000A 2% | |
| 5.001A-40.000A 3% | |
| DCR Test | ● |
| Constant power output | ● |
| Current switch | ● |
6355/6356 LCR Meter
| Model | 6355/6356 | |
| Frequency Range | 10Hz-200kHz/10Hz-500kHz | |
| Frequency Setting | Continuity | |
| Frequency Output Accuracy | ±0.01% | |
| Frequency Resolution | 5-bit Frequency | |
| Basic Accuracy | ±0.05% | |
| Drive Level | Test Signal Voltage Level | AC:10mV – 2Vrms DC:10mV – 2V |
| Voltage Minimum Resolution | 1mV | |
| Output Impedance | 100Ω | |
| Measurement Mode | Meter mode, list mode | |
| Measurement Parameters | |Z|、|Y|、θ、X、R、G、B、L、D、Q、DCR、C | |
| Test Time (Fastest) | 8mS | |
6355/6356 Measurement Parameters
| Parameters | Range |
| |Z|, R, X | 0.001mΩ~99.999MΩ |
| |Y|, G, B | 0.1nS~99.999S |
| Cs Cp | 0.01pF~9.9999F |
| Ls Lp | 0.1nH~99.999kH |
| D | 0.00001~9.9999 |
| Q | 0.1~9999.9 |
| DCR | 0.001mΩ~99.999MΩ |
| Δ% | -9999%~999% |
| θ | -180° ~+180° |
System
| Measurement Circuit | Series/Parallel |
| Correction | Open Circuit/Short Circuit |
| List Mode | 8 steps |
| Bulit-in Storge | 128 sets |
| USB Host Storge | LCR setting documents, list mode setting document |
| Trigger Test | Auto, manual, RS-232, GPIB, Handler |
| Power Supply | Voltage:88~132Vac or 195~264Vac (Switchable) |
| Frequency:47-63Hz | |
| Power consumption | 300VA (6223)、600VA (6243H/6243) |
| Environment | Temperature:10℃-40℃ |
| Humidity:20-90%RH | |
| Dimension | 358x148x343 mm (W*H*D) /6355 |
| 337×145×525mm (W*H*D) /6223 | |
| 435×145×644mm (W*H*D) /6243H/6243 | |
| Weight | 3Kg (6355/6356) |
| 15Kg (6223) | |
| 20Kg (6243H/6243H/6243) | |
| Display | Color Screen, 7″ TFT (800*480)|6355 |
| Interface | RS-232、Handler |