The CS Q-LEAP™ system is based on the HERO™ vibration controller and the SE-201 shock exciter. It enables primary and secondary calibration of shock-sensors as well as complete measuring instruments with very high precision and efficiency, according to ISO 16063 – 13, ISO 16063-22 and ISO 17025.
Key features
- Broad amplitude range: 5 gn…10 000 gn (49 m/s²…98 km/s²)
- Traceable to PTB (German National Metrology Laboratory)
- Primary calibration of shock sensors
- Integrated sensor database
- Integrated e-Cal™ software for the generation of calibration certificates (print, PDF…)
- Easy data exchange with applications like ERP systems or measuring equipment databases
Technical Data
- Broad amplitude range: 5 gn…10 000 gn (49 m/s²…98 km/s²)
- Pulse width: 0.1 ms…5 ms
- Automated regulation of amplitudes: up to 600 gn (6 km/s²)
- Max. payload: 80 g (2.82 oz)
CS Q-LEAP™ software
- shock calibration
- vibration measurement
- vibration generation
- more on demand
Typical DUTs
- vibration/shock sensors
· PE transducers
· IEPE transducers
· PR transducers
· Digital transducers (SPI, I2C, DTI and other interfaces) - supports TEDS / ID modules according to IEEE 1451.4
Standards
- ISO 16063 – 13: Primary shock calibration using laser interferometry
- ISO 16063 – 22: Calibration of vibration transducers by comparison to a reference transducer
- ISO 17025: General requirements for the competence of testing and calibration laboratories