Primary vibration calibration
The CS Q-LEAP™ system is used for primary vibration calibration of a wide range of sensors including vibration sensors and measurement systems. It is based on the HERO™ vibration controller and the SE-09 high-frequency vibration exciter.
Key features
- Frequency range: 5 Hz…20 kHz (50 kHz)
- Traceable to PTB (German National Metrology Laboratory)
- Calibration of vibration sensors, measurement systems and calibrators
- Integrated sensor database
- Integrated e-Cal™ software for the generation of calibration certificates (print, PDF…)
- Easy data exchange with applications like ERP systems or measuring equipment databases
Technical Data
- Max. acceleration: 400 m/s² (40 gn) peak
- Max. velocity: 0.5 m/s (19.7 inch)
- Max. displacement: 7.5 mm (0.3 inch)
- Max. payload: 350 g (12 oz)
CS Q-LEAP™ Software
- sine calibration
- sine sweep
- vibration measurement
- vibration generation
- more on demand
Typical DUTs
- PE transducer
- IEPE transducer
- VC transducer
- PR transducer
- digital transducers with SPI, I2C, DTI, and many other interfaces
- vibration meter
- laser vibrometer
- vibration calibrator
Standards
- ISO 16063 -11: Primary calibration of vibration transdurcers
- ISO 16063 – 21: Calibration of vibration transducers by comparison to a reference transducer
- ISO 16063 – 41: Calibration of laser vibrometers
- ISO 17025: General requirements for the competence of testing and calibration laboratories